Section 4 User Help Reference
4-19
Insert Point — This command or button will insert a
row below the selected point.
Delete Point — This command will delete the selected
row. Note that you cannot delete the initial point.
PeakTrak automatically updates the Run Length setting or
scales the other points when you change the number of points
and their duration on the curve.
To close the Gradient Table, click the Edit Gradient button.
See Also: D
EFINING A GRADIENT [4.4.5]
4.3.2.5 Peak Collection
Use the Peak Collection buttons to set the collection mode.
All — Click this button to collect all fluids during a run.
Peaks — Click this button to collect only eluted peaks
during a run.
None — Click this button to divert all fluids to the
waste port during a run.
Tube Volume — You can also specify the Tube Volume
for collected fluids. This volume can be the default
maximum volume (Max option) as entered in the
C
ONFIGURATION [4.3.10] window, or a method-specific
volume less than the capacity defined by the
C
ONFIGURATION [4.3.10] window. When using a Foxy
fraction collector or the Fractionation Valve, note that
the actual fraction size may be less if a newly detected
peak causes a tube change, or if you click the Next Tube
button.
Peak — Type or select the desired volume to be
collected in each tube when the module detects a
peak.
Non-peak — Type or select the desired volume to be
collected in each tube when peaks are not detected.
This setting allows you to conserve tubes without
diverting non-peak elute to waste. This setting is
ignored when the peak collection mode is set to Peaks
or None.