Section 4 User Help Reference
4-17
Time Windows can limit the fraction collection to specific
time durations of the run. To use time windows, first
enter a start time and an end time. Note that all times
are relative to start of the run, just following the sample
injection. Then choose a peak collection mode. If you
select Peaks the module will collect fluid only when the
time window is active and a peak is present. If you select
All the module will collect all fluid during the time
window regardless of the peak state. Define up to three
time windows.
Air Purge — Type or select the duration of a post-run
air purge, in minutes. An air purge clears the solvent
from the column and the fluid path, forcing any liquid
out the waste port. If you are disposing the column after
each run, a sufficient post-run air purge ensures that it
will be empty. To disable the air purge, enter a value of 0
(zero). A column can also be manually air purged. See
M
ANUAL CONTROL [4.3.9].
4.3.2.3 Run Notes
Use this text entry box to enter comments or notes for the run.
These comments will be saved with the run and will appear in
TXT and PDF reports.
4.3.2.4 Gradient Plot Area
The gradient plot area depicts the current gradient profile
that will be used for the run. The X-axis of the plot displays
the run time in minutes or C
OLUMN VOLUMES [ON-LINE]. The
X-axis scale can be adjusted by typing or selecting the Run
Length. The left Y-axis displays absorbance units and the
right Y-axis indicates the percentage of solvent B.
The gradient plot area also provides a convenient method to
edit the gradient curve. You can click and drag any of the
points to change the shape of the curve, or use the buttons at
the bottom of the plot area to insert and delete points. The
following controls appear in the Gradient Plot Area:
Insert — Click this button to enable the gradient point
insert mode. When this mode is active, click the gradient
curve to add a single point. You can then drag the new
point to any desired position. Click the insert button